- Item
- Drywall Temperature Calibrator
- Includes
9142 Field Metrology Well, Insert for Configuration Ordered, Power Cord, RS-232 Cable, User Guide, Technical Manual CD, Report of Calibration, Calibration Label, 6-pin DIN Connector, Test Lead Kit, Well Insulator, Clamp-on ferrites (3), Tongs (insert removal tool), 9930 Interface-it Software and User's Guide
- Display
- LCD
- Standards
- NVLAP Accredited
- Interface
- RS-232
- Loop Supply Voltage
- 24V DC
- Accuracy
- +/-0.01 Degrees C
- RTD Type
- NI-120, 385, 3892, JIS
- Application
- Field Testing
- RTD Temperature Range
- -196 to 660 Degrees C
- Thermocouple Type
- C, E, J, K, L, M, N, R, S, T, U
- Max. Loop Current
- 24 mA
- Thermocouple Temperature Range
- 0 to 1480 Degrees C (type S)
- Calibration Certificate
- NVLAP accredited
- Agency Compliance
- CE, CSA, C-TICK, ROHS, WEEE
- Resistance Sensing Current Characteristics
- 0.02% at 2 mV; Cal 4-22 mA, Spec 4-24 mA
- Temperature Range
- -196 Degrees to 660 Degrees C, 0 Degrees to 1,480 Degrees C
- Reference Input Temperature Characteristics
- +/-0.010 Degrees C at -25 Degrees C, +/-0.015 Degrees C at 0 Degrees C, +/-0.020 Degrees C at 50 Degrees C, +/-0.025 Degrees C at 150 Degrees C, +/-0.030 Degrees C at 200 Degrees C, +/-0.040 Degrees C at 350 Degrees C, +/-0.050 Degrees C at 420 Degrees C, +/-0.070 Degrees C at 660 Degrees C
- RTD Temperature Range C
- -196 Degrees to 660 Degrees C
- Typical Cooling Time
- 15 min (23 Degrees to -25 Degrees C), 25 min (150 Degrees to -23 Degrees C)
- Stability
- +/-0.01 Degrees C
- Thermocouple mV Characteristics
- 0.025% + 0.01 mV at -10 to 75 mV
- Calibrator Type
- Field Metrology Well
- Typical Heating Time
- 16 min (23 Degrees to 140 Degrees C), 23 min (23 Degrees to 150 Degrees C), 25 min (-25 Degrees to 150 Degrees C)
- Sensor Type
- RTD, Thermocouple
- Thermocouple Temperature Range C
- 0 Degrees to 1,480 Degrees C
- Display Resolution
- 0.01 Degrees F
- Reference Input Resistance Characteristics
- +/--0.0025 ohm Accuracy (0 to 42 ohm), +/-60 ppm of Reading Accuracy (42 to 400 ohm)
- Current mA Characteristics
- 0.02% + 2 mV Accuracy/4 to 22 mA, 4 to 24 mA Range
- Probe Connection Types
- (1) 4-Terminal RTD Input, (1) Small Connectors TC
- Brand and Series
- Fluke 9142
- Includes NIST-Traceable Certificate
- No
- Power Source Details
- 100 to 115V, 230V
- For RTD Type
- 4-Wire, Ni-120
- Reference Junction Temperature Characteristics
- +/- 0.35 Degrees C Accuracy at Ambient Temperature of 13 Degrees to 33 Degrees C
- Calibrator Application
- Industrial Process Environment
- Thermocouple Temperature Accuracy Characteristics
Type C: +/-1.1 Degrees C at 660 Degrees C, Type E: +/-0.7 Degrees C at 660 Degrees C, Type J: +/-0.7 Degrees C at 660 Degrees C, Type K: +/-0.8 Degrees C at 660 Degrees C, Type L: +/-0.7 Degrees C at 660 Degrees C, Type M: +/-1.4 Degrees C at 660 Degrees C, Type N: +/-0.9 Degrees C at 660 Degrees C, Type R: +/-1.4 Degrees C at 660 Degrees C, Type S: +/-1.5 Degrees C at 660 Degrees C, Type T: +/-0.8 Degrees C at 400 Degrees C, Type U: +/-0.75 Degrees C at 600 Degrees C
- Calibrator Accuracy
- +/-0.2 Degrees C (Display - Full Range)
- For Thermocouple Type
- Type C, Type E, Type J, Type K, Type L, Type M, Type N, Type R, Type S, Type T, Type U
- Calibrator Weight
- 18 lb
- Typical Uniformity
- +/-0.05 Degrees C, +/-0.01 Degrees C
- With Process Connections
- Yes
- Display Temperature Accuracy
- +/-0.2 Degrees C
- Single Well/Dual Well
- Single Well
- Number of Set Points
- Variable
- Set Point Style
- Adjustable
- 4-Wire Device Resistance Characteristics
- Range: 0 to 400 ohms/Accuracy: +/-0.0025 ohms (0 to 42 ohms), +/-60 ppm of Reading (42 to 400 ohms)
- Temperature Accuracy
- +/-0.01 Degrees C
- Calibration Details
- NVLAP Accredited Calibration Included
- Sensor Technology
- RTD, Thermocouple
- RTD Temperature Celsius - Maximum
- 660 Degrees C
- Includes Process Connections
- Yes
- Thermocouple Temperature Celsius - Minimum
- 0 Degrees C
- Thermocouple Temperature Celsius - Maximum
- 1,480 Degrees C
- Voltage
- 100 to 115V, 230V
- Process Connections Included
- Reference PRT, RTD, Thermocouple, Transmitter
- RTD Temperature Celsius - Minimum
- -196 Degrees C
- Series
- 9142