Drywell Temperature Calibrator, 9140 Series, Field Dry-Well, 95 to 662 Degrees F, Single Well
Mfr # 9140-A-156
Zoro # G8928692
product price:Â $4,745.00
/ea
Free Shipping & Returns
Key Features
Accuracy
+/-0.5 Degrees C
Interface
RS-232
Calibration Certificate
NIST Traceable
Display
LED
Agency Compliance
CE, ROHS, WEEE
Get a 10% off coupon
Free 30-day returns Return Policy
Get 30 days to pay with Net 30 Apply Now
Unlock tax-exempt savings Add Tax Exemption
Specifications
Product Information
MFR #
9140-A-156
Zoro #
G8928692
UPC #
095969128926
Country of Origin
United States
Dimensions
Overall Dimensions
3 3/8 in H x 6 in W x 7 3/4 in D
For Stem Diameter
1/16 in, 1/8 in, 3/16 in, 1/4 in, 3/8 in, 1/2 in
Overall Width
6 in
Overall Depth
7.75 in
Overall Height
3.38 in
Description
The FLUKE 9140-A-156 is a field dry-well temperature calibrator used for checking and calibrating temperature probes and thermometers across a range of 35 Degrees C to 350 Degrees C (95 Degrees F to 662 Degrees F). Designed for high-volume field applications, it features a single well with adjustable set points, LED display, and RS-232 interface for communication and software use. Its compact size and 6 lb weight support portability for on-site temperature calibration work.
Key Features of the FLUKE 9140-A-156 Drywell:
- Temperature Range: Operates from 35 Degrees C to 350 Degrees C (95 Degrees F to 662 Degrees F) for a broad range of calibration points.
- Accuracy: Rated at +/-0.5 Degrees C, with calibrator accuracy of +/-0.25 Degrees C and +/-1 Degrees C for holes greater than 1/4 in.
- Stability and Uniformity: Stability is +/-0.03 Degrees C at 50 Degrees C and +/-0.05 Degrees C at 350 Degrees C, with typical uniformity of +/-0.1 Degrees C.
- Display and Resolution: LED display with 0.1 Degrees resolution for temperature reading and set point visibility.
- Probe Sizes: Supports stem diameters of 1/16 in, 1/8 in, 3/16 in, 1/4 in, 3/8 in, and 1/2 in.
- Heating and Cooling Times: Typical heating time is 12 minutes from ambient to 350 Degrees C; typical cooling time is 15 minutes from 350 Degrees C to 100 Degrees C.
- Stabilization Time: Reaches stabilization in 7 minutes.
- Included Items: Supplied with the 9140 dry-well, insert for the ordered configuration, power cord, insert removal tool, 9930 Interface-it software, serial cable, and user's guide.
- Calibration Documentation: Includes a NIST traceable calibration certificate with data at 50 Degrees C, 100 Degrees C, 150 Degrees C, 200 Degrees C, 250 Degrees C, 300 Degrees C, and 350 Degrees C.
- Size and Power: Measures 3 3/8 in H x 6 in W x 7 3/4 in D and operates on 115V AC or 230V AC.
- Standards: CE, ROHS, and WEEE.