Item
Drywall Temperature Calibrator
Includes
6 ft Main Power Cord, 9190 Insert D, 9930 Interface-It Calibration Software and Users Guide, Getting Started Manual, Insert Removal Tool, Product CD, Report of Calibration and Calibration Label, USB Cable, Well Insulator Cap
Standards
IEC 61010-1, Installation Category II, Pollution degree 2, IEC 61326-1: Basic
Interface
RS-232, USB Cable
Loop Supply Voltage
24V DC
Accuracy
+/-0.2 Degrees C Full Range
RTD Type
4 Wire, 3 Wire, 2 Wire
Application
Blood Banks, Cold Rooms, Freeze Dryers, Laboratory Refrigerators, Medical Freezers, Sterilizers (Autoclaves)
RTD Temperature Range
-139 Degrees to 284 Degrees F (-95 Degrees to 140 Degrees C)
Thermocouple Type
C, E, J, K, L, M, N, R, S, T, U
Max. Loop Current
4 to 20mA
Agency Compliance
EU Directives, CSA, Australian EMC Requirements
Thermocouple Temperature Range
-139 Degrees to 284 Degrees F (-95 Degrees to 140 Degrees C)
Calibration Certificate
yes
Display Temperature Accuracy
+/-0.2 Degrees C
Calibrator Type
Field Metrology Well
Thermocouple Temperature Range C
-95 Degrees to 140 Degrees C
Typical Cooling Time
60 min (140 Degrees to 23 Degrees C), 80 min (23 Degrees to -90 Degrees C), 90 min (23 Degrees to -95 Degrees C)
Current mA Characteristics
0.02% + 0.002 mA Accuracy/4 to 22 mA, 4 to 24 mA Range
RTD Temperature Range C
-95 Degrees to 140 Degrees C
RTD Temperature Range F
-139 Degrees to 284 Degrees F
Stability
+/-0.015 Degrees C
Reference Input Temperature Characteristics
+/-0.010 Degrees C at -95 Degrees C, +/-0.013 Degrees C at -25 Degrees C, +/-0.015 Degrees C at 0 Degrees C, +/-0.020 Degrees C at 50 Degrees C, +/-0.025 Degrees C at 140 Degrees C
Stabilization Time
10 min
Typical Heating Time
40 min (-95 Degrees to 140 Degrees C)
Power Source Details
100 to 115V, 200 to 230V
Sensor Type
RTD, Thermocouple
For Thermocouple Type
Type C, Type E, Type J, Type K, Type L, Type M, Type N, Type R, Type S, Type T, Type U
Brand and Series
Fluke 9190A
For RTD Type
2-Wire, 3-Wire, 4-Wire, Ni-120, PT-100 (385), PT-100 (3926), PT-100 (JIS)
Reference Junction Temperature Characteristics
+/- 0.35 Degrees C Accuracy at Ambient Temperature of 13 Degrees to 33 Degrees C
Reference Input Resistance Characteristics
+/--0.0025 ohm Accuracy (0 to 42 ohm), +/-60 ppm of Reading Accuracy (42 to 400 ohm)
Probe Connection Types
(1) 4-Terminal RTD Input, (1) Miniature Connectors (ASTM E1684)
Thermocouple Temperature Range F
-139 Degrees to 284 Degrees F
Resistance Sensing Current Characteristics
0.02% at 0.002 mA; Cal 4-22 mA, Spec 4-24 mA
Thermocouple mV Characteristics
0.025% + 0.01 mV at -10 to 75 mV
With Process Connections
Yes
Display Resolution
0.01 Degrees F
Thermocouple Temperature Accuracy Characteristics
Type C: +/-1.00 Degrees C at 140 Degrees C, Type E: +/-0.60 Degrees C at 140 Degrees C, Type J: +/-0.70 Degrees C at 140 Degrees C, Type K: +/-0.75 Degrees C at 140 Degrees C, Type L: +/-0.65 Degrees C at 140 Degrees C, Type M: +/-0.65 Degrees C at 140 Degrees C, Type N: +/-0.75 Degrees C at 140 Degrees C, Type R: +/-1.60 Degrees C at 140 Degrees C, Type S: +/-1.60 Degrees C at 140 Degrees C, Type T: +/-0.60 Degrees C at 140 Degrees C, Type U: +/-0.70 Degrees C at 140 Degrees C
Calibrator Accuracy
+/-0.2 Degrees C
Typical Uniformity
+/-0.05 Degrees C
Temperature Range
-139 Degrees to 284 Degrees F/-95 Degrees to 140 Degrees C
4-Wire Device Resistance Characteristics
Range: 0 to 400 ohms/Accuracy: +/-0.0025 ohms (0 to 42 ohms), +/-60 ppm of Reading (42 to 400 ohms)
Number of Set Points
Variable
Single Well/Dual Well
Single Well
Set Point Style
Adjustable
Temperature Accuracy
+/-0.2 Degrees C
Temperature Set Points
-95 Degrees C, 140 Degrees C
Process Connections Included
Reference PRT, RTD, Thermocouple, Transmitter
Interchangeable Insert
Yes
RTD Temperature Fahrenheit - Maximum
284 Degrees F
RTD Temperature Celsius - Maximum
140 Degrees C
Thermocouple Temperature Celsius - Maximum
140 Degrees C
Includes Process Connections
Yes
Calibration Certificate Type
Accredited
Thermocouple Temperature Fahrenheit - Minimum
-139 Degrees F
Number of Inserts Included
1
Thermocouple Temperature Celsius - Minimum
-95 Degrees C
RTD Temperature Celsius - Minimum
-95 Degrees C
RTD Temperature Fahrenheit - Minimum
-139 Degrees F
Sensor Technology
RTD, Thermocouple
Thermocouple Temperature Fahrenheit - Maximum
284 Degrees F
Voltage
100 to 115V, 200 to 230V